Comet T3417 Specifications |
Technical Parameters |
Value |
Output |
RS485 |
Measured Value |
Temperature + Relative Humidity |
Construction Type |
Duct Mount (700mm) |
Design |
Industrial |
Relative humidity range |
0 to 100% |
Accuracy of relative humidity measurement |
±2.5% relative humidity from 5 to 95% at 23°C |
Accuracy of temperature output |
±0.4°C |
Available temperature units |
degrees Celsius, Fahrenheit |
Accuracy and range of dew point temperature output - for more details see graphs |
±1.5°C at ambient temperature T<25°C and RH>30% range -60 to +80 °C |
Accuracy and range of absolute humidity output |
±1.5g/m3 at ambient temperature T < 25°C range 0 to 400 g/m3 |
Accuracy and range of specific humidity output |
±2g/kg at ambient temperature T < 35°C range 0 to 550 g/kg |
Accuracy and range of mixing ratio output |
±2g/kg at ambient temperature T < 35°C range 0 to 995 g/kg |
Accuracy and range of specific enthalpy output |
±3kJ/kg at ambient temperature T < 25°C range: 0 to 995 kJ/kg |
Temperature operating range |
-30 to +80°C |
Temperature operating range of LCD display |
readable to operating temperature +70°C, it is recommended to switch OFF the LCD over +70°C |
Temperature compensation of the humidity sensor |
all temperature range |
Communication protocol |
ModBus RTU and Advantech ADAM compatible protocol |
Communication speed |
110 to 115200 Bd |
Filtering ability of sensor cover |
0.025mm - filter with stainless steel mesh |
Protection of the case with electronics |
IP65 electronics, IP40 sensors |
Power |
9-30Vdc, current consumption approximately 0.5W |
Dimensions |
display unit 88.5 x 94 x 39.5 mm (W x H x D), stem length 700 mm |
Weight |
approximately 580g |