High accuracy electric fiber-insulated chamber furnace line consists of 4 models. These universal laboratory furnaces are designed for materials testing, heat treatment such as hardening, loosening, normalising, ceramic and stoneware samples firing up to 1300°C. To eliminate gasses or smokes that are released during thermal processing, exhaust system may be additionally installed in the products. The furnaces are excellent fit for scientific laboratories, educational institutions, ceramic studios, medicine and industry.
Description
-
Maximum temperatures 1100°C, 1200°C, 1300°C,
-
Capacity 30L, 40L, 80L,
-
Chamber made of fiber thermal insulation blocs,
-
Vacuumized heating elements in four sides around a chamber (1100°C),
-
Heating elements in grooves (1200°C),
-
Heating elements on tubes (1300°C),
-
Outside casing in metal sheet, powder painting grey,
-
Door opens to the right side,
-
Door safety interlock switch,
-
Microprocessor temperature controller (Omron E5CC – basic model),
-
Ceramic bottom plates,
-
Fast heating time due to low thermal mass construction,
-
Low power consumption,
-
Good stability and uniformity.
Technical data |
Dimensions |
SNOL 80/1100 LSF01 |
Useful volume |
Liter |
80 |
Rated power not more then |
kW |
5,4 |
Rated supply voltage |
V |
400 |
Rated frequency |
Hz |
50 |
Number of phases |
- |
1 |
Continuous operating temperature |
°C |
50-1100 |
Maximum temperature |
°C |
1100 |
Working chamber material |
- |
Fiber plates |
Maximum heating- up time (without charge) |
Min. |
150 |
Temperature stability in working chamber at rated temperature in thermal steady state without charge not more than |
± °C |
2 |
Temperature uniformity in working space at rated temperature in thermal steady state without charge not more than |
±°C |
10 |
Furnace working chamber dimensions: |
|
|
width |
mm |
300 |
depth |
mm |
450 |
height |
mm |
600 |
Furnace dimensions: |
|
|
width |
mm |
740 |
depth |
mm |
880 |
height |
mm |
1230 |
Mass ( Netto) |
kg |
135 |
Control
-
Temperature measurement by thermocouple type ”K”/”S”
-
PID electronic regulator, double digital display reference temperature and measured temperature.
-
Products are equipped with high-precision digital microprocessor Omron or Eurotherm temperature controllers fitted with self-tuning and manual PID settings. The customer can select a basic or programmable temperature controller, which offers up to 32 programming segments (rate of temperature rise or decrease control, maintenance of preset temperature, automatic shutdown). A wide range of devices allows selecting the most appropriate controller for your process.
-
SSR control unit.
Supplied Time
-
Furnace and temperature controller instructions,
-
Electric diagram,
-
CE declaration.
Production Time
-
Basic model usually on stock
Guarantee
-
One year limited guarantee and later service for furnace.
Options
-
Buzzer,
-
Timer,
-
OTP (over temperature protection),
-
Chimney with ventilator for aching process,
-
Eurotherm 3216 (non-programmable),
-
Eurotherm 3208 (programmable),
-
Omron E5CC-T (programmable),
-
PC connection and computer software SNOL V2.0,
-
Metal tray (box) for samples treatment,
-
Protective gas box & equipment for gas inlet.
Country of Origin
-
Lithuania (EU country).
-
CE marked.
Payment Condition
Packing
Spare Parts
Fiber thermoinsulation line, € |
80/1100 |
Chamber (with heating elements) |
|
Thermocouple |
|
Ceramic bottom plates set |
|